Please send all questions to product@genscript.com.
Problem
|
Possible Cause(s)
|
Solution(s)
|
---|---|---|
Little or no amplification product
|
GC rich template
|
add 5% DMSO, 1 M betaine, or both can be included in PCR reactions to improve results
|
Template degradation
|
Analyze the template on an agarose gel to check for possible degradation |
|
PCR inhibitor contamination
|
|
|
Cycle conditions
|
|
|
Primer problem
|
|
|
Low enzyme concentration
|
Increase the amount of polymerase in 0.5 U steps
|
|
Low MgCl2 concentration
|
Increase the MgCl2 concentration in 0.25mM steps
|
|
Inadequate annealing temperature
|
Increase annealing temperature
|
|
Improper template concentration
|
Test serial dilution of template
|
|
Formation of primer dimers
|
When the temperature of the PCR cycle reaches 94°C insert PCR reaction tube and continue the program
|